Télécharger Imprimer la page

Keysight U1253B Guide D'utilisation Et De Maintenance page 175

Masquer les pouces Voir aussi pour U1253B:

Publicité

6
Performance Tests and Calibration
Table 6-2
Performance verification tests
Step
Test function
1
Turn the rotary switch to the
2
Press
to switch to frequency mode
3
Press
to switch to duty cycle mode
4
Turn the rotary switch to the
Press
to select DC V measurement
174
Range
5 V
[a]
position
50 V
500 V
1000 V
9.9999 kHz
0.01% to 99.99% 5.0 Vpp @ 50%,
5 V
position
50 V
500 V
1000 V
Reference signals/
values
5520A output
5 V, 1 kHz
5 V, 10 kHz
4.5 V, 20 kHz
4.5 V, 30 kHz
4.5 V, 100 kHz
50 V, 1 kHz
50 V, 10 kHz
45 V, 20 kHz
45 V, 30 kHz
45 V, 100 kHz
500 V, 1 kHz
1000 V, 1 kHz
0.48 V, 1 kHz
square wave, 50 Hz
5 V
50 V
500 V
1000 V
Keysight U1253B User's and Service Guide
Error limits
± 22.5 mV
± 79.0 mV
± 0.1695 V
± 0.1695 V
± 0.1695 V
± 225.0 mV
± 790.0 mV
± 1.695 V
± 1.695 V
± 1.695 V
± 2.25 V
± 8.0 V
± 500 mHz
± 0.315%
± 1.75 mV
± 17.5 mV
± 200 mV
± 800 mV

Publicité

loading