6
Performance Tests and Calibration
Table 6-3
Valid adjustment reference input values (continued)
Function
Range
AC mA/AC A
50 mA
500 mA
5 A
10 A
Capacitance
Open
10 nF
100 nF
1000 nF
10 F
100 F
1000 F
10 mF
190
Reference input value
3.000 mA (1 kHz)
30.000 mA (1 kHz)
30.00 mA (1 kHz)
30.000 mA (1 kHz)
0.3000 A (1 kHz)
3.0000 A (1 kHz)
0.3000 A (1 kHz)
10.000 A (1 kHz)
OPEN
3.000 nF
10.000 nF
10.00 nF
100.00 nF
100.0 nF
1000.0 nF
10.000 F
100.00 F
1000.0 F
10.000 mF
Valid range for reference input
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
Open terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
Keysight U1253B User's and Service Guide