Optimizing Throughout and Reading Rate
A Benchmark
The benchmark used to show the affect of the various functions of the 3458A
multimeter will start with the most convenient, but least rapid, procedure of
having the computer ask the dmm to change to a particular function, make a
measurement, and transfer the measurement to the computer. The benchmark
will assume that all of the measurements will be made through a FET scanner of
infinitely fast switching speed and of infinite dynamic range. Hence, the
benchmark represents an artificial situation, but one where the different modes of
operation of the 3458A can be best illustrated. The computer used is the 9000
Series 200/300. Times for other computers will vary depending on the GPIB
turnaround time of the computer. Results are shown in
The DUT contains:
25 resistance measurements
15 < 10 kOhm ± 5%
8 < 100 kOhm ± 5%
2 < 10 kOhm ± .001%
10 DCV measurements
5 < 30 V± 1%
4 < 10 V± .01 %
1 < 1 V± .001%
3 ACV measurements
1 < 250 V @ 50 Hz @ ± 5%
1 < 10 V @ 25 kHz @ ± 0.1%
1 < 1 V @ 5 kHz @ ± 0.075%