EXTOUT
External output. Specifies the event that will generate a signal on the rear panel
Ext Out connector (EXTOUT signal). This command also specifies the polarity of
the EXTOUT signal.
Syntax
EXTOUT [event][,polarity]
event
The event choices are:
event
parameter
OFF
ICOMP
ONCE
APER
BCOMP
SRQ
RCOMP
Power-on event = ICOMP.
Default event = ICOMP.
264
Numeric
Description
query equiv.
0
None; EXTOUT is disabled
Input complete (1 µs pulse after A/D converter has integrated
1
each reading or, for direct- or sub-sampling, after the track and
hold has acquired the input signal)
Outputs a 1 µs pulse upon execution of the EXTOUT ONCE
2
command; the event then becomes OFF
Aperture waveform (a level indicating when the A/D converter is
3
making a measurement)
4
Burst complete (1 µs pulse following a group of readings)
Status event occurred (1 µs pulse whenever a status register event
5
occurs that has been enabled to assert the GPIB SRQ). (See
second Remark below.)
6
Reading complete (1 µs pulse after each reading)