Figure 4-6
Using an external scanner
Burst complete
When specified, the burst complete event (BCOMP event) produces a 1 µs pulse
following completion of a group of readings. The number of readings in a group is
specified by the NRDGS or SWEEP command. The BCOMP event can be used to
synchronize an external scanner to the multimeter when making multiple readings
per scanner channel. The following program is similar to the preceding program
except that it uses the BCOMP event and makes 15 readings on each scanner
channel. Connections for this example are shown in
10 OUTPUT 722;"PRESET NORM"!DCV, NRDGS 1,AUTO, TARM AUTO, TRIG SYN
20 OUTPUT 722;"MEM FIFO"!ENABLE READING MEMORY, FIFO MODE
30 OUTPUT 722;"TRlG EXT"!TRIGGER EVENT = EXTERNAL
40 OUTPUT 722;"EXTOUT BCOMP, NEG"!BURST COMPLETE EVENT, LOW-GOING TTL
50 OUTPUT 722;"NRDGS 15, AUTO"!15 READINGS PER CHANNEL
55!CONFIGURE EXTERNAL SCANNER
60 OUTPUT 709;"SADV EXTIN"!ADVANCE SCANNER ON MULTIMETER'S EXTOUT
SIGNAL
70 OUTPUT 709;"CHCLOSED EXT"!OUTPUT LOW-GOING PULSE AFTER EACH CLOSURE
80 OUTPUT 709;"SCAN 201- 206"!SCAN CHANNELS 01 - 06 ON SCANNER IN SLOT
200