Télécharger Imprimer la page

Keysight Technologies 3458A Guide D'utilisation page 332

Multimètre

Publicité

6
Command Reference
– Although RESET can be used from remote, it is intended primarily for front
– When attempting to send the RESET command from remote, it is possible that
– Related commands: PRESET
Example
10 CLEAR 722
20 OUTPUT 722;"RESET"
30 END
332
ALL math registers set to 0 except:
DEGREE = 20
SCALE
= 1
PERC
= 1
panel use. RESET configures the multimeter to a good starting point for local
operation. Executing the RESET command from the alphabetic menu resets
the multimeter as shown above. Pressing the shifted front panel Reset key,
however, has the same effect as cycling the multimeter's power. This stores
the present state as state 0, any compressed subprograms are destroyed,
stored readings are destroyed, the power-on SRQ bit is set in the status
register, and the power-on sequence is performed.
the multimeter is busy or the GPIB bus is being held. In either case, the
multimeter will not respond immediately to the remote RESET command. For
this reason, you should send the GPIB device clear command before you send
the multimeter's RESET command. This is shown in the example below.
REF = 1
RES = 50
!CLEARS THE MULTIMETER IMMEDIATELY
!RESETS THE MULTIMETER
Keysight 3458A User's Guide