6
Command Reference
TERM
On previous multimeters, the TERM command internally connected or
disconnected the multimeter's input terminals. The 3458A accepts the TERM
command to maintain language compatibility with these multimeters, but does
not respond since the 3458A's input terminals cannot be controlled from remote.
Syntax
TERM [source]
source
The source parameter choices are:
source
parameter
OPEN
FRONT
REAR
Power-on source = none.
Default source = FRONT.
Remarks
– Query command: The TERM? query command returns a response indicating
370
Numeric
query equiv.
0
1
2
which input terminals (FRONT or REAR) are selected by the front panel
Terminals switch.
Description
Generates error message
Generates error message if Terminals switch is set to Rear
Generates error message if Terminals switch is set to Front
Keysight 3458A User's Guide