operation
parameter
OFF
CONT
CTHRM
DB
DBM
FILTER
FTHRM
NULL
PERC
PFAIL
RMS
SCALE
STAT
CTHRM2K
CTHRM10K
Keysight 3458A User's Guide
Numeric
Description
Equiv.
0
Disables all post-process math operations
Enables the previous math operation. To resume two math
1
operations, send MMATH CONT,CONT
Result = temperature (Celsius) of a 5 kW thermistor (40653B).
3
Function must be OHM or OHMF (10 kW range or higher).
Result = 20 x log
4
initialized to 1, yielding dBV.
Result = 10 x log
5
be ACV, DCV, or ACDCV.
Result = output of exponentially weighted digital low-pass filter.
6
Response is set by DEGREE register.
Result = temperature (Fahrenheit) of a 5 kW thermistor (40653B).
8
Function must be OHM or OHMF (10 kW) range or higher).
Result = reading-OFFSET register. The OFFSET register is set to
9
first reading—after that you can change it.
10
Result = ((reading - PERC register) / PERC register) x 100.
11
Reading vs. MAX and MIN registers.
Result = squares reading, applies FILTER operation, takes square
12
root.
13
Result = (reading-OFFSET register) / SCALE register.
Performs statistical calculations on the present set of readings
and stores results in these registers:
SDEV = standard deviation
14
MEAN = average of readings
NSAMP = number of readings
UPPER = largest reading
LOWER = smallest reading
Result = temperature (Celsius) of a 2 kW thermistor (40653A).
16
Function must be OHM or OHMF.
Result = temperature (Celsius) of a 10 kW thermistor (40653C).
17
Function must be OHM or OHMF.
Command Reference
(reading/REF register). The REF register is
10
2
(reading
/RES register/1 mW). Function must
10
6
295