– Related commands: MSIZE, PURGE, SCRATCH, SSTATE
Example
OUTPUT 722; "RSTATE B2" !RECALLS STORED STATE NAMED B2
SCAL
This is a calibration command. Refer to the 3458A Calibration Manual for details.
SCRATCH
Clears all subprograms and stored states from memory.
Syntax
SCRATCH
Remarks
– Individual subprograms can be cleared with the DELSUB command. Individual
– Related commands: DELSUB, PURGE, RSTATE, SSTATE, SUB
Example
OUTPUT 722;"SCRATCH" !CLEARS ALL SUBPROGRAMS AND STORED STATES
SECURE
Security code. Allows the person responsible for calibration to enter a security
code to prevent accidental or unauthorized calibration or autocalibration
(autocal). (Refer to the
Syntax
SECURE old_code, new_code[,acal_secure]
old_code
This is the multimeter's previous security code. The multimeter is shipped from
the factory with its security code set to 3458A.
Keysight 3458A User's Guide
states can be cleared with the PURGE command.
ACAL
command for details on autocal.)