Table 3-9
The following program configures the multimeter for frequency measurements on
the 10 V range from a voltage Source. The input signal is AC-coupled.
10 OUTPUT 722;"FREQ 10"
20 OUTPUT 722;"FSOURCE ACV"
30 END
The following program configures the multimeter for period measurements on the
10 mA range from a current source. The input signal is DC-coupled.
10 OUTPUT 722;"PER 10E-3"
20 OUTPUT 722;"FSOURCE ACDCI"
30 END
You can reduce high-frequency noise above 75 kHz for frequency or period
NOTE
measurements by enabling the level filter. Refer to the LFILTER command in
Chapter 6
Specifying bandwidth
The ACBAND command specifies the frequency content of the input signal for all
AC and AC+DC measurements. Specifying the frequency content allows the
multimeter to make accurate measurements and to configure itself for the fastest
possible measurements. The ACBAND command's first parameter specifies the
lowest expected frequency component (the power-on/PRESET NORM value is
20 Hz). The second parameter specifies the highest expected frequency
Keysight 3458A User's Guide
FSOURCE parameters
FSOURCE
parameter
ACV
AC-coupled AC voltage input
ACDCV
DC-coupled AC voltage input
ACI
AC-coupled AC current input
ACDCI
DC-coupled AC current input
for details.
Definition
Configuring for Measurements
Measurement capabilities
Frequency
1 Hz — 10 MHz
1 Hz — 10 MHz
1 Hz — 100 kHz
1 Hz — 100 kHz
3
Period
100 ns — 1 s
100 ns — 1 s
10 µs — 1 s
10 µs — 1 s
103