Command Reference
– When you execute the RMEM command from the front panel, readings are
– In addition to the RMEM command, you can also recall readings using the
– Related commands: MCOUNT?, MEM, MFORMAT, MSIZE, NRDGS
Example
10 OUTPUT 722;"TARM HOLD" !SUSPENDS TRIGGERING
20 OUTPUT 722;"DCV" !DC VOLTAGE MEASUREMENTS
30 OUTPUT 722)"TRIG AUTO" !AUTOMATIC TRIGGERING
40 OUTPUT 722;"NRDGS 3 ,AUTO" !3 READINGS PER SAMPLE EVENT (AUTO)
50 OUTPUT 722;"MEM FIFO" !ENABLES READING MEMORY, FIFO MODE
60 OUTPUT 722;"TARM SGL, 10" !10 GROUPS OF READINGS
70 OUTPUT 722;"RMEM 1,3,6" !READS 1ST - 3RD READINGS of 6TH GROUP
80 ENTER 722;A,B,C !ENTERS READINGS INTO A, B, & C VARIABLES
90 PRINT A,B,C !PRINTS READINGS
100 END
336
has the highest number. Numbers are always assigned in this manner
regardless of whether you're using the FIFO or LIFO mode. Records are also
numbered in this manner—the most recent record is record number 1.
copied, one at a time, to the display. After viewing the first reading, you can
view others by using the up or down arrow key. Use the left and right arrow
keys to view the reading number (left side of display) and the reading (right
side of display).
"implied read." Refer to
Recalling readings
in
Chapter 4
for more information.