E
E
5. Insert Sealed Dispersion Tool
As delivered, the drive unit is equipped for the standard dispersion tool (non-
sealed). To use the sealed dispersion tool, the drive unit must be converted.
To convert the drive unit, place it in a horizontal position. Then insert the screwdriver (Q) (in-
cluded) into the slit in the inner ring of the ball bearing 1 (R) and unscrew it clockwise. Remove
the ball bearing 1 (R) for use with the standard dispersion tool. Then return the drive unit to an
upright position.
Insert the sealed dispersion tool (N) into the hole (O) in the drive unit (U) up to the stop and then
screw it on by rotating it counterclockwise. Secure with set screw (P).
The following rules apply to the conversion of the drive unit:
For non-sealed dispersion tools, insert ball bearing 1 (R).
For sealed dispersion tools, remove ball bearing 1 (R).
Q
R
O
P
N
Q
U
- 36 -
R