5 Hygienic preparation

5
Hygienic preparation

5.1 General information

5.2 Intervals

5.3 Hygienic preparation of the module

Requirement
Part
Cleaning
Housing
Wipe with a damp cloth
using water or mild soap
Connection line
36
EN
prisma2CLOUD
All manuals and user guides at all-guides.com
Wear suitable protective equipment for disinfection work.
Please refer to the instructions for use supplied with the
disinfectant used.
Interval
Weekly
As necessary, e.g., following infectious
diseases or severe soiling
Damage to the device caused by ingress of liquids!
Ingress of liquids may damage the therapy device, components,
and accessories.
 Do not immerse the therapy device, components, or
accessories in liquids.
The therapy device is switched off (see instructions for use for the
therapy device).
1. Remove the module from the therapy device
Removing the module", page
2. Carry out hygienic preparation of the module as specified in
the following table:
Action
Clean the modules
Disinfect the modules
35).
Disinfection
Wipe-down disinfection
(Recommendation: terralin
protect or perform advanced
Alcohol EP)
(see "4.2
Sterilization
®
Not permitted