16
5. Hygienic preparation
Keep the STATION MEDUMAT clean by
simply wiping it with disinfectant.
Observe the instructions for the disinfectant
used.
We recommend Terralin.
You are recommended to wear suitable
gloves (e.g. household or disposable gloves)
during disinfection procedures.
6. Operational inspection
Wall bracket STATION MEDUMAT
WM 22550 and STATION MEDUMAT with
bracket WM 22950 must be subjected to an
operational inspection before each use, but at
least every 6 months.
6.1 Test interlock
1. Insert a MEDUMAT Standard/Basic in
STATION MEDUMAT and allow the in-
terlock to engage.
2. Check that the MEDUMAT is securely
supported in the STATION MEDUMAT,
without opening the interlock.
MEDUMAT and STATION MEDUMAT
must now be rigidly connected and must
not be allowed to be separated.
English