a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the ME Equipment is used exceeds the applicable RF compliance level
above, the ME Equipment should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the ME Equipment.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
TEST SPECIFICATIONS FOR ENCLOSURE PORT IMMUNITY TO RF WIRELESS COMMUNICATION EQUIPMENT
BAND (A)
TEST FREQUENCY
(MHZ)
385
380-395
450
430-470
710
745
704-787
780
810
870
800-960
930
1 720
1 845
1 700-1 990
1 970
2 450
2 400-2 570
5 240
5 500
5 100-5 800
5 785
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be
reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
(a) For some services, only the uplink frequencies are included.
(b) The carrier shall be modulated use a 50% duty cycle square wave signal.
(c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be
worst case.
WARRANTY
Masimo warrants to the initial buyer only that these products, when used in accordance with the directions
provided with the Products by Masimo, will be free of defects in materials and workmanship for a period of six
(6) months. Single use products are warranted for single patient use only.
THE FOREGOING IS THE SOLE AND EXCLUSIVE WARRANTY APPLICABLE TO THE PRODUCTS SOLD BY MASIMO
TO BUYER. MASIMO EXPRESSLY DISCLAIMS ALL OTHER ORAL, EXPRESS OR IMPLIED WARRANTIES, INCLUDING
WITHOUT LIMITATION ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE.
MASIMO'S SOLE OBLIGATION AND BUYER'S EXCLUSIVE REMEDY FOR BREACH OF ANY WARRANTY SHALL BE, AT
MASIMO'S OPTION, TO REPAIR OR REPLACE THE PRODUCT.

WARRANTY EXCLUSIONS

This warranty does not extend to any product that has been used in violation of the operating instructions
supplied with the product, or has been subject to misuse, neglect, accident or externally created damage. This
warranty does not extend to any product that has been connected to any unintended instrument or system, has
been modified, or has been disassembled or reassembled. This warranty does not extend to sensors or patient
cables that have been reprocessed, reconditioned or recycled.
IN NO EVENT SHALL MASIMO BE LIABLE TO BUYER OR ANY OTHER PERSON FOR ANY INCIDENTAL, INDIRECT,
SPECIAL OR CONSEQUENTIAL DAMAGES (INCLUDING WITHOUT LIMITATION LOST PROFITS), EVEN IF ADVISED
OF THE POSSIBILITY THEREOF. IN NO EVENT SHALL MASIMO'S LIABILITY ARISING FROM ANY PRODUCTS SOLD
TO BUYER (UNDER A CONTRACT, WARRANTY, TORT OR OTHER CLAIM) EXCEED THE AMOUNT PAID BY BUYER
FOR THE LOT OF PRODUCT(S) INVOLVED IN SUCH CLAIM. IN NO EVENT SHALL MASIMO BE LIABLE FOR ANY
DAMAGES ASSOCIATED A PRODUCT THAT HAS BEEN REPROCESSED, RECONDITIONED OR RECYCLED. THE
LIMITATIONS IN THIS SECTION SHALL NOT BE DEEMED TO PRECLUDE ANY LIABILITY THAT, UNDER APPLICABLE
PRODUCTS LIABILITY LAW, CANNOT LEGALLY BE PRECLUDED BY CONTRACT.
NO IMPLIED LICENSE
This single-patient sensor is licensed to you under the patents owned by Masimo for single-patient use only. By
acceptance or use of this product, you acknowledge and agree that no license is granted for use of this product
with more than a single patient.
After single-patient use, discard sensor.
Purchase or possession of this sensor confers no express or implied license to use the sensor with any device
which is not separately authorized to use Masimo sensors.
SERVICE (A)
TETRA 400
GMRS 460, FRS 460
LTE Band 13, 17
GSM 800/900, TETRA 800,
iDEN 820, CDMA 850, LTE
Band 5
GSM 1800; CDMA 1900; GSM
1900; DECT; LTE Band 1, 3. 4.
35: UMTS
Bluetooth, WLAN, 802.11
b/g/n, RFID 2450, LTE Band 7
WLAN 802.11 a/n
MODULATION (B)
Pulse modulation (b) 18 Hz
FM (c) +/- 5 kHz deviation 1 kHz
sine
Pulse modulation (b)
217 Hz
Pulse modulation (b)
18 Hz
Pulse modulation (b)
217 Hz
Pulse modulation (b)
217 Hz
Pulse modulation (b)
217 Hz
7
MAXIMUM
DISTANCE
POWER (W)
(M)
1,8
0,3
2
0,3
0,2
0,3
2
0,3
2
0,3
2
0,3
0,2
0,3
10714E-eIFU-0122