RECOMMENDED SEPARATION DISTANCES
The ME Equipment is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
the ME Equipment can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the ME Equipment as recommended below, according to the maximum output power of the communication equipment.
RATED MAXIMUM OUTPUT POWER OF TRANSMITTER
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
The ME Equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the ME Equipment should assure
that it is used in such an environment.
EMISSION TEST
RF Emissions
CISPR 11
RF Emissions
CISPR 11
The ME Equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the ME Equipment should
assure that it is used in such an environment.
IMMUNITY TEST
Electrostatic
+/- 8 kV contact
discharge (ESD)
+/- 15 kV air
IEC 61000-4-2
Power frequency
(50 / 60 Hz)
30 A/m
magnetic field.
IEC 61000-4-8
Portable and mobile RF communications equipment should be used no closer to any part of the ME Equipment, including cables, than the
recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
IMMUNITY TEST
10 V/m
Radiated RF
80 MHz to
IEC 61000-4-3
2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
RECOMMENDED SEPARATION DISTANCE BETWEEN PORTABLE AND
MOBILE RF COMMUNICATION EQUIPMENT AND THE ME EQUIPMENT
80 MHz to 800 MHz
d = 1.17*√P
0.12
0.37
1.17
3.7
11.7
GUIDANCE AND MANUFACTURER'S DECLARATION- ELECTROMAGNETIC EMISSIONS
COMPLIANCE
The ME Equipment must emit electromagnetic energy in order to perform its intended function.
Group 1
Nearby electronic equipment may be affected.
Class B
Suitable for use in all establishments, including domestic environments.
GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
IEC 60601 TEST
COMPLIANCE LEVEL
LEVEL
+/- 8 kV contact
+/- 15 kV air
30 A/m
IEC 60601 TEST
COMPLIANCE LEVEL
LEVEL
10 V/m
SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER (M)
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Guidance -
Power frequency magnetic fields should be at levels characteristic of typical
location in a typical hospital environment.
RECOMMENDED SEPARATION DISTANCE
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey a, should be less than the compliance
level in each frequency range b.
Interference may occur in the vicinity of equipment marked with
the following symbol:
6
800 MHz a 2.5 GHz
d = 2.33*√P
0.23
0.74
2.33
7.37
23.3