• Do not attempt to sterilize by irradiation, steam, autoclave or ethylene oxide as it will damage the sensor.
• Do not attempt to reuse on multiple patients, reprocess, recondition or recycle Masimo sensors or patient
cables as these processes may damage the electrical components, potentially leading to patient harm.
• Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the Radius T°, including cables
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
• Keep the Radius T° away from electrical equipment that emits radio frequencies to minimize radio
interference. Radio interference may result in no or inaccurate readings.
• The frequency bands of this device (2.4 GHz) are only for indoor use, in accordance with international
telecommunication requirements.
INSTRUCTIONS
A) Site Selection
• Select a site on the left side of the chest where the skin is clean of debris and dry prior to sensor
placement. Refer to Fig. 1.
• The site should be hair-free, cleaned of debris and dry prior to sensor placement.
B) Applying the sensor
1. Open the package and remove the sensor.
2. Pull and remove the plastic battery tab. Refer to Fig. 2.
3. Peel off the release liner from the sensor. Refer to Fig. 3.
Note: Avoid contact with the exposed sensor adhesive.
4. Place the sensor to the selected application site.
Note: Ensure that the skin of the patient is relaxed and not stretched in any way and that there are no skin
folds under the sensor pad.
5. Press around the perimeter of the sensor to ensure the adhesive is secure to the patient's skin.
C) Pairing the Sensor
1. Once battery tab is removed, sensor is available for Bluetooth pairing. Refer to Fig. 4.
2. A solid blue light indicates the sensor is connected.
3. Check the application display to ensure the sensor is communicating correctly.
Note: Readings may take up to 15 minutes to appear on the application.
4. Periodically check the sensor or application for a solid blue light to confirm that is it connected.
D) Sensor Reapplication
Note: Radius T° sensors are designed for removal and reapplication no more than one (1) time over the
life of the product.
1. Clean and dry the sensor application site.
2. Gently wipe the exposed sensor adhesive with an alcohol wipe and allow to dry to restore the adhesive
properties.
3. Follow steps 1 through 5 from above to re-apply the sensor.
E) Removing the Sensor
1. Peel gently to remove the sensor from the patient.
Note: Disposal of Product: Comply with local laws in the disposal of the sensor, battery and its accessories.
LIGHT INDICATOR GUIDE/TROUBLESHOOTING
Color
Sensor
No light
---
Green
flashing
flashing
Blue
solid
Orange
flashing
Red
flashing
For additional help, contact Masimo Technical Services at (949) 297-7498. Local contact information can be found at:
http://service.masimo.com.
Description
• Sensor power is off.
• Sensor is on and waiting to pair with host device.
• Sensor is waiting for user confirmation that desired sensor was
paired to the host device.
• Successful pairing of sensor and host device.
• Host device successfully receiving data.
• Low sensor battery
• Depleted sensor battery
• Hardware or sensor failure, sensor blinking board failure code
• Confirm battery pull tab has been removed to
• Replace the sensor.
• Follow instructions to pair with the host device.
• Verify sensor attachment so that host device
• Consider replacing the sensor.
• Replace the sensor.
4
Next steps
activate the battery.
can receive data.
10714E-eIFU-0122