Troubleshooting; Technical Data; Operation Conditions; Transport And Storage Con- Ditions - Precision Tech Lumina Smart Manuel Utilisateur

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

9. TROUBLESHOOTING

When trouble is found, check the following points before contacting your distributor. If none of these are
applicable or the trouble is not remedied even after action has been taken, the product may have failed.
Contact your distributor.
PROBLEM
The power is not turned
on.
The Power LED does not
light up when charging.
The handpiece does
not display the charging
interface while charging.
No output.
No sound.
Insufficient light
intensity.

10. TECHNICAL DATA

Manufacturer
Model
Dimensions
Weight
Power supply
Charger power supply
Charger power output
Power Frequency
Charger nominal power input
Light intensity
Wavelength
Electrical safety class
Applied part
34
CAUSE
The battery is run down.
The time to press the main switch is
too short.
Use a wrong adapter.
There is no electricity in the outlet.
The adapter is not connected.
Put the handpiece into the charge
base in the wrong direction.
The charge base broken.
Charge pin of charge base unable to
rebound.
Handpiece broken.
Beep volume set to off.
Handpiece broken.
There are resin or other contaminants
on the surface of the lamp lens.
Changzhou Sifary Medical Technology Co.
Lumina Smart
22cm x 11cmx 8cm±1cm
730g ±10%
Lithium ion battery: 3.7V, 1600mAh, ±10%
AC 100-240 V, ±10%
5V
1 A
50/60Hz, ±10%
5VA
2300mW/cm
- 1500mW/cm
2
- 600 mW/cm
2
385nm-515nm
Class II
B
SOLUTION
Charge the battery.
Press the main switch more than
0.5 seconds.
Use the original adapter.
Check the connection.
Check the connection.
Check the direction.
Using adapter connect to
handpiece directly, and contact
your distributor.
Remove debris which between
move part and base of the
charge pin.
Contact your distributor.
Set beep volume to low, mid or
high.
Contact your distributor.
Cleaning the lamp head residue.

OPERATION CONDITIONS

Use: in enclosed spaces
Ambient temperature: 5°C ~ 40 °C
Relative humidity: <80%
Operating altitude < 3000m above sea level
Atmospheric pressure: 70kPa ~ 106kPa
TRANSPORT AND STORAGE CON-
- 1000mW/cm
2
2
DITIONS
Ambient temperature: -20 °C ~ +55 °C
Relative humidity: 20% ~ 80 %
Atmospheric pressure: 70kPa ~ 106kPa
REF. CHAP.
6.1
4.1
/
/
/
3.4
/
/
/
/