TEST
BAND
FREQUENCY
(MHz)
(MHz)
1720 - 1845 -
1700-1990
1970
2450
2400-
2570
5240, 5500,
5100-
5785
5800
12. STATEMENT
Service Life
The service life of Lumina Smart products is 3 years.
Disposal
The package should be recycled. Metal parts of the device are disposed as scrap metal. Synthetic
materials, electrical components, and printed circuit boards are disposed as electrical scrap. The lithium
batteries are disposed as special refuse. Please deal with them according to the local environmental
protection laws and regulation.
Rights
All rights of modifying the product are reserved to the manufacturer without further notice. The pictures
are only for reference. The final interpretation rights belong to CHANGZHOU SIFARY MEDICAL
TECHNOLOGY CO., LTD. The industrial design, inner structure, etc, have claimed for several patents by
SIFARY, any copy or fake product must take legal responsibilities.
Warranty
Lumina Smart is covered by 1 year warranty.
SERVICE
MODULATION
GSM 1800,
Pulse
CDMA 1900,
modulation 217 Hz
GSM 1900,
DECT, LTE
Band 1, 3, 4,
25; UMTS
Bluetooth,
Pulse
WLAN,
modulation 217 Hz
802.11 b/g/n,
RFID 2450,
LTE Band 7
WLAN
Pulse
802.11 a/n
modulation 217 Hz
MAXIMUM
DISTANCE
POWER
(m)
(W)
2
0.3
2
0.3
0.2
0.3
IMMUNITY
TEST LEVEL
(V/m)
28
28
9
37