GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The Lumina Smart is intended for use in the electromagnetic environment specified below. The customer or the user
of the Lumina Smart should assure that it is used in such an environment.
Immunity test
Conducted dis-turbances
induced by RF fields
IEC 61000-4-6
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
wireless communication
equipment
IEC 61000-4-3
RECOMMENDED MINIMUM SEPARATION DISTANCES
Nowadays, many RF wireless equipments have being used in various healthcare locations where medical equipment
and/or systems are used. When they are used in close proximity to medical equipment and/or systems, the medical
equipment and/or systems' basic safety and essential performance may be affected. The Lumina Smart has been
tested with the immunity test level in the below table and meet the related requirements of IEC 60601-1-2:2014. The
customer and/or user should help keep a minimum distance between RF wireless communications equipments and
the Lumina Smart as recommended below.
TEST
BAND
FREQUENCY
(MHz)
(MHz)
385
380-390
450
430-470
710 - 745-780
704-787
810 - 810 -
800-960
930
36
IEC 60601 test level
3 V
0.15 MHz – 80 MHz, 6
V in ISM bands be-tween
0.15 MHz and 80 MHz,
80 % AM at 1 kHz
3 V/m, 80 MHz – 2,7 GHz,
80 % AM at 1 kHz
See the RF wireless
communication equipment
table in "Recommended
minimum separation
distances"
SERVICE
MODULATION
TETRA 400
Pulse modulation
GMRS 460
FRS 460
± 5 kHz deviation 1
LTE Band
13, 17
modulation 217 Hz
GSM
800/900,
modulation 18 Hz
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Compliance
Electromagnetic environment - guidance
level
3 V
Portable and mobile RF communications
equipment should be usedno closer to
any part of the Lumina Smart, including
cables, than the recommended separation
distance calculated from the equation ap-
plicable to the frequency of the transmitter.
Recommended minimum separation
3 V/m
distances
See the RF wireless communication equi-
pment table in "Recommended minimum
separation distances"
Complies
MAXIMUM
POWER
(W)
1.8
18 Hz
FM
2
kHz sine
Pulse
0.2
Pulse
2
IMMUNITY
DISTANCE
TEST LEVEL
(m)
(V/m)
0.3
27
0.3
28
0.3
9
0.3
28