Specification
Circuit Test
Meter reading range:
Open Circuit Voltage:
Short Circuit Current:
Fuse Check
Earth Continuity Test
Open Circuit Voltage:
Constant Current:
10A Earth Bond Test
Meter reading range:
Pass Band Limits:
Open Circuit Voltage:
0 to 9,9 kΩ
Range
Resolution
0 to 1 kΩ
0,01 kΩ
1 kΩ to 2 kΩ
0,02 kΩ
2 kΩ to 5 kΩ
0,05 kΩ
5 kΩ to 9,9 kΩ
0,1 kΩ
4,5 V d.c.
1 mA d.c.
5 V. d.c. 0,5 mA Typical
Range
Resolution
0 to 999 mΩ
1 mΩ
900 mΩ to1990 mΩ
10 mΩ
Typically 100 mV d.c.
Typically 100 mA into <1Ω
0 to 1990 mΩ
Range
Resolution
0 to 600 mΩ
1 mΩ
600 to 999 mΩ
1 mΩ
900-1990 mΩ
10 mΩ
100 and 500 mΩ
Typically 3 V rms a.c. 50 Hz
Accuracy
2,5% of reading ± 0,02 kΩ
2,5% of reading ± 0,02 kΩ
2,5% of reading ± 0,05 kΩ
2,5% of reading ± 0,1 kΩ
Accuracy
2,5% of reading ±5 mΩ
mΩ
2,5% of reading ±10
Accuracy
2,5% of reading ±5 mΩ
Current below level for
satisfactory Bond test.
Transfer to 25A test
15