Proceed as follows:
1. Tighten the threaded plug into the Foot Pedal Socket.
2. Then, push the pedal hard repeatedly to detect possible breaks in the tubing or in the pedal.
D. If the Electrosurgical Generator, correctly connected, appears to deliver a lower power output than
usual, check:
1. The Dispersive Patient Plate for complete contact (refer to Dispersive Plate sections).
2. The condition of active electrodes (refer to "Section 10, Practical Suggestions").
3. The condition of the handpiece (cable continuity, contact of the electrode in the handpiece and
the connector) by moving the cable and the connector and electrode to detect possible breaks
and poor contact in the socket of the system.
E. The Electrosurgical Generator has a thermal protection circuit which will shut it off when internal
operating temperatures exceed safe limits. Should the Electrosurgical Generator stop functioning
with no alarm signal, ensure that the system has adequate ventilation and that you have not exceeded
the recommended duty cycle of 10/30 seconds. Should the thermal protection circuit be activated
under normal operating conditions, the Electrosurgical Generator must be returned to CooperSurgical
for servicing (see Section 15).
Section 14 Accessories
NOTE: Use only genuine CooperSurgical KH1000A accessories (patient return pads, hand pencils,
electrodes and smoke evacuation disposables) for optimal system performance and patient safety. The
use of non-CooperSurgical authorized and genuine accessories is not recommended nor have been
tested and verified to the level of safety and performance of a genuine CooperSurgical accessory.
Contact CooperSurgical for a list of current genuine KH1000A accessories.
Section 15 Liability Statement
CooperSurgical guarantees the safety, reliability and performance of the LEEP System 1000
only if the installation, recalibrations and repairs are performed by personnel authorized by CooperSurgical
and if it is used in compliance with given instructions in an area that meets all the applicable IEC
requirements.
Section 16 Warranty
CooperSurgical, Inc., warrants that the LEEP System 1000 Workstation (the "Product") will be free from
defects in materials and workmanship for a period of one (1) year from the original date of purchase.
If the product should become inoperable due to a defect in material or workmanship during this one-year
warranty period, CooperSurgical will, at its option, repair or replace the product. This limited warranty does
not include replacement or service to repair damage resulting from improper installation, external electrical
fault, accident, disaster, use for a purpose other than that for which originally designed or indicated in this
manual, negligence, modification, service or repair by personnel not authorized by CooperSurgical or normal
wear and tear, and also does not apply to disposable or single- or limited-use items or components. The sole
and exclusive remedy under this limited warranty shall be repair or replacement as provided herein.
The foregoing limited warranty states the sole warranty made by CooperSurgical with respect to the product
and all parts thereof, and is in lieu of any other warranty by CooperSurgical with respect to product.
COOPERSURGICAL NEITHER MAKES NOR GRANTS ANY OTHER WARRANTY, EITHER EXPRESS
OR IMPLIED, WITH RESPECT TO THE PRODUCT, INCLUDING WITHOUT LIMITATION, WARRANTIES
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT WILL
COOPERSURGICAL BE LIABLE FOR ANY DAMAGES ARISING OUT OF THE LOSS OF USE OF
THE PRODUCT, OR ANY OTHER INCIDENTAL OR CONSEQUENTIAL DAMAGES, WHETHER OR NOT
COOPERSURGICAL HAS ADVANCE KNOWLEDGE OF THE POSSIBILITY OF SAME.
No person, agent, distributor, dealer or company is authorized to change or modify the terms of this
limited warranty.
®
LEEP System 1000
Workstation • Model KH1000A • Operating Manual • English
®
(continued)
Workstation
23