Electromagnetic Compatibility Information

Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic emissions
The SBC 24 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the SBC 24 should assure that it is used in
such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fl uctuations / fl icker emissions
IEC 61000-3-3
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The SBC 24 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the SBC 24 should assure that it is used in
such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency (50 / 60 Hz)
magnetic fi eld IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
All manuals and user guides at all-guides.com
Compliance
Group 1
Class B
Not applicable
Not applicable
IEC 60601 test level
Compliance level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic environment – guidance
The SBC 24 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
The SBC 24 is suitable for use in all establishments other than domestic and those
directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If fl oors are covered with
synthetic material, the relative humidity should be at least 30 %.
Power frequency magnetic fi elds should be at levels characteristic of a typi-
cal location in a typical commercial or hospital environment.