Electromagnetic Compatibility Information

Table 1
Guidance and manufacturer's declaration – electromagnetic emissions
The SBC 25 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the SBC 25 should assure
that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fl uctuations/fl icker emissions
IEC 61000-3-3
Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
The SBC 25 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the SBC 25 should assure
that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency (50/60 Hz)
magnetic fi eld IEC 61000-4-8
All manuals and user guides at all-guides.com
Compliance
Electromagnetic environment – guidance
Group 1
The SBC 25 uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equip-
ment.
Class B
The SBC 25 is suitable for use in all establishments, including domestic establish-
ments and those directly connected to the public low-voltage power supply network
Not applicable
that supplies buildings used for domestic purposes.
Not applicable
IEC 60601 test level Compliance level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If fl oors are
covered with synthetic material, the relative humidity should be
at least 30 %.
Power frequency magnetic fi elds should be at levels charac-
teristic of a typical location in a typical commercial or hospital
environment.