Electromagnetic Compatibility Information

Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration – electromagnetic emissions
The SBC 29 is intended for use in the electromagnetic environment specified below. The customer of the user of the SBC 29 should assure that it is
used in such and environment.
Emission test
Compliance
RF emissions CISPR 11
RF emission CISPR 11
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The SBC 29 is intended for use in the electromagnetic environment specified below. The customer or the user of SBC 29 should assure that it is used
in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency (50Hz) magne-
tic field
IEC 61000-4-8
Electromagnetic environment – guidance
Group 1
The SBC 29 uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
Class B
The SBC 29 is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buil-
dings used for domestic purposes.
IEC 60601 test level
± 6 kV contact
± 8 kV air
3 A/m
Compliance level
Electromagnetic environment - guidance
± 6 kV contact
Floors should be wood, concrete or ceramic
± 8 kV air
tile. If floor are covered with synthetic material,
the relative humidity should be at least 30%.
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
65