Electromagnetic Compatibility Information

For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration – electromagnetic emission
The SBC 15 is intended for use in the electromagnetic environment specified below.
The customer or the user of the SBC 15 should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration – electromagnetic immunity
The SBC 15 is intended for use in the electromagnetic environment specified below. The customer or the user of the SBC 15 should assure that it is
used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Compliance
Electromagnetic environment – guidance
Group 2
The SBC 15 must emit electromagnetic energy in order to perform its intended function.
Nearby electronic equipment may be affected.
Class B
The SBC 15 is suitable for use in all establishments other than domestic and those directly
connected to the public low-voltage power supply network that supplies buildings used for
Not applicable
domestic purposes.
Not applicable
IEC 60601 test
Compliance
level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with syn-
thetic material, the relative humidity should be at least 30 %.
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.