Technical speciications
Operation temperature/
humidity:
Storage temperature/
humidity:
Weight:
Dimensions:
Measuring method:
Measurement range:
Cuff pressure display:
Voltage source:
40
•
10 to 40 °C (50 to 104 °F)
•
15 - 90 % relative maximum humidity
•
-20 to 55 °C (-4 to 131 °F)
•
15 - 90 % relative maximum humidity
•
1100 g (including rechargeable
battery pack)
•
200 x 125 x 90 mm
•
Oscillometric, corresponding to
Korotkoff
•
30 - 280 mmHg – blood pressure
•
40 - 200 beats per minute – pulse
•
Range: 0 - 299 mmHg
•
Resolution: 1 mmHg
•
Static accuracy: pressure within ± 3
mmHg
•
Pulse accuracy: ±5 % of the readout
value
•
Rechargeable battery pack;
4.8V C3500 mAh
•
Mains adapter DC 7.5V, 2 A
Reference to Standards: • Device corresponds to the
requirements of the standard for non-
invasive blood pressure monitor.
EN 1060-1
EN 1060-3
EN 1060-4
IEC 60601-1
IEC 60601-1-2
Electromagnetic
• Device fulills the stipulations of the
standard IEC 60601-1-2.
compatibility:
The stipulations of the EU Directive 93/42/
EEC for Medical Devices Class IIa have been
fulilled.
Type BF applied
part
Microlife reserves the right to alter technical
specications without prior written notice.