Special Functions

Hiding measurement values

The WatchBP Ofice ABI device features a Hide
function in order to prevent unnecessary elevated
blood pressure in patients due to nervousness
triggered by visible blood pressure values.
This feature is available in both SCREEN mode and
*
ROUTINE mode.
When the «Hide» function is ON, only the average
*
values are displayed after the consecutive
measurements and the Hide icon is displayed.
1) Activate the «Hide» Function – Press M button;
L or R lashes; keep pressing and holding the M
Button for 7 more seconds until On or Off lashes.
L or R lashes...
2) Select ON or OFF – Press the M Button again to
turn the "Hide" function ON or OFF.
3) Conirm – Press Start/Stop to conirm the setting.
The default setting of "Hide" is set at "ON" in both
*
«SCREEN» Mode and «ROUTINE».
Press and hold
for 7 seconds...
25