3.4 CH1 / CH2
Tracking Parallel
Mode
Background /
connection
Output rating
Panel operation
DC Power Supply
Tracking parallel operation doubles the current capacity of the power supply
by internally connecting CH1 and CH2 in parallel and combining the output
to a single channel. CH1 controls the combined output.
0 - 30 V / 0 ≈ 6 A
1. Press the PAR/INDEP key to activate the
tracking parallel mode. The key LED turns on.
2. Connect the load to the CH1 +/- terminals.
3. To turn on the output, press the output key.
The key LED turns on.
4. The CH2 indicator turns red, indicating tracking
parallel (PAR) mode
5. Press the CH1 switch (LED turns on) and use
the Voltage and Current knob to set the output
voltage and current. The CH2 output control is
disabled. By default, the Voltage and Current
knob work in the coarse mode. To activate the
fine mode, press the knob and turn on the
FINE LED.
6. For the output level and CV/CC status, refer to
the CH1 meter and indicator.
Voltage level: The CH1 meter reading shows
the output voltage. In the above case, 20.0 V.
Current level: Double the amount of CH1
current meter reading. In the above case,
2.0 A x 2 = 4.0 A
IV - 45