Operation (contd.)
IV - 44
4. Press the CH1 switch (LED turns on) and use
the Voltage knob to set the master & slave
output voltage (the same level for both
channels). By default, the Voltage and
Current knob work in the coarse mode. To
activate the fine mode, press the knob and
turn on the FINE LED.
Coarse: 1V or 0.1A @ rotation click.
Fine: the smallest digit @ rotation click.
4. Use the current knob to set the master output
current.
5. To turn on the output, press the output key.
The key LED turns on.
6. For the master (CH1) output level and CV/CC
status, refer to the CH1 meter and indicator.
Master (CH1) voltage level: CH1 meter
reading shows the output voltage. In the
above case, 20.0 V.
Master (CH1) current level: CH1 meter
reading shows the output current. In the
above case, 2.000 A.
7. Press the CH2 switch (LED turns on) and use
the Current knob to set the slave output
current.
8. For the slave (CH2) output level and CV/CC
status, refer to the CH1/2 meter and CH2
indicator.
Master (CH1) voltage level: CH1 meter
reading shows the output voltage. In the
above case, 20.0 V.
Master (CH1) current level: CH1 meter
reading shows the output current. In the
above case, 2.000 A.