Setup

Power Up

Connect AC power
cord
Power on
Power off
Load Cable
Connection
Banana plug
Wire type
III - 38
This chapter describes how to properly power up the power supply before
operation.
Connect the AC power cord to the rear
panel socket.
Press the power switch to turn on the power. The display shows the
initialization screen with the model name, followed by the last recalled
settings.
Press the power switch again to turn off
Insert the ∅ 4 mm safety
When using load cables other than the attached, make sure they have
enough current capacity for minimizing cable loss and load line impedance.
Voltage drop across a wire should not excess 0.5 V. The following list is the
wire current rating at 450 A/c.
Wire size (AWG)
20
18
16
14
12
the power.
banana plug.
Max. current (A)
2.5
4
6
10