Operation (contd.)
3. CH1/CH2 Tracking
Serial Mode
3.1 Tracking serial
without common
terminal
Connection
Output rating
Panel operation
IV - 42
Tracking series operation doubles the Voltage capacity of the power supply
series by internally connecting CH1 (Master) and CH2 (Slave) in serial and
combining the output to a single channel. CH1 (Master) controls the
combined Voltage output level.
The following describes two types of configurations depending on the
common ground usage.
0 - 60 V / 0 - 3 A
1. Press the SER/INDEP key to activate the
tracking serial mode. The key LED turns on.
2. Connect the load to the front panel terminals,
CH1+ & CH2-. (Single supply).
3. Press the CH2 switch (LED turns on) and use
the Current knob to set the CH2 output current
to the maximum level. By default, the Voltage
and Current knob work in the coarse mode. To
activate the fine mode, press the knob and turn
on the FINE LED.
Coarse: 1V or 0.1A @ rotation click.
Fine: the smallest digit @ rotation click.
4. Press the CH1 switch (LED turns on) and use
the Voltage and Current knob to set the output
voltage and current level..
Operation
DC Power Supply