Cleaning
WARNINGS:
▪
ALWAYS clean the equipment as indicated upon initial receipt and
before each use.
▪
DO NOT resterilize or reuse the needles, syringe, or catheter.
Discard these components after one use.
CAUTIONS:
▪
DO NOT immerse any component in liquid. DO NOT allow liquids
or moisture to enter any electrical connection.
▪
DO NOT sterilize the Pressure Monitor.
▪
DO NOT use solvents, lubricants, or other chemicals, including
glutaraldehyde or similar chemical cleaners, unless otherwise
specified.
▪
DO NOT use unapproved disinfectants. Failure to comply may
cause equipment damage.
Recommended Equipment and Materials
▪
PPE as recommended by the disinfectant supplier
▪
Soft, lint-free cloth
▪
United States Environmental Protection Agency (US EPA)
registered disinfectant with a claim for activity against Hepatitis
B. The following disinfectants have been validated for use on the
exterior surfaces of the equipment:
Sodium Hypochlorite Based - Clorox® Clean-Up® Disinfectant
Cleaner with Bleach (EPA Reg. #67619-17)
Quaternary Ammonium Based - CaviCide® (EPA Reg. #46781-6)
To Wipe Down the Equipment
1. Turn off the monitor.
2. Wipe all external surfaces of the equipment with a soft, lint-
free cloth moistened with a non-abrasive, hospital disinfectant
prepared according to the manufacturer's instructions.
3. Upon the removal of visible gross soil, use a clean cloth
moistened with disinfectant and wipe all surfaces. Surfaces
must remain visibly wet at room temperature for at least the
minimum time specified in the instructions for use supplied by the
disinfectant manufacturer.
4. Remove any excess disinfectant solution using a soft, lint-free
cloth moistened with water if required by the instructions supplied
by the disinfectant manufacturer.
5. Inspect the equipment.
8
Testing and Maintenance
WARNINGS:
▪
Only individuals trained and experienced in the maintenance
of this reusable medical device should test and maintain this
equipment.
▪
DO NOT attempt to remove the screws on the bottom of the
monitor and open the device. The electronics are coated for
moisture resistance and are not serviceable.
NOTE: For service, contact your C2Dx sales representative or call
C2Dx customer service. Outside the US, contact your nearest C2Dx
subsidiary.
Battery Replacement
1. Turn off the monitor.
2. Remove the back cover of the monitor with a paper clip to
expose the battery.
3. Pull the battery out of the unit.
4. Unsnap the battery from the connector.
5. Replace the battery with a new, 9 V alkaline battery.
6. Tuck the battery and the wires back inside the monitor.
7. Replace the cover, and snap the cover into place.
Calibration Verification
Verify the calibration of the Pressure Monitor (REF 0295-001-000) per
hospital protocol or at least once per year.
CAUTION: Do not attempt to repair this unit. If recalibration is
necessary, return the Pressure Monitor to C2Dx.
There are two methods of calibration verification: The Water Method
(with a disposable unit), and the Air Method (without a disposable
unit).
NOTE: The Water Method is preferred because it accounts for
preloading characteristics of the monitor.
0295-001-072 Rev-1