EN
The Intra-Compartmental Pressure Monitor System (REF 0295-001-000) is intended for use in the electromagnetic environment specified below. The customer or the
user of the Intra-Compartmental Pressure Monitor System (REF 0295-001-000) should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE: U
is the alternating current mains voltage prior to application of the test level.
T
Recommended separation distances between portable and mobile RF communications equipment and the Intra-Compartmental Pressure Monitor System
The Intra-Compartmental Pressure Monitor System (REF 0295-001-000) is intended for use in the electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Intra-Compartmental Pressure Monitor System (REF 0295-001-000) can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Intra-Compartmental Pressure Monitor System
(REF 0295-001-000) as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
12
Guidance and manufacturer's declaration - electromagnetic immunity
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3 A/m
(REF 0295-001-000)
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
,
n/a
n/a
n/a
n/a
n/a
Compliance level
±2, 4, 6 kV contact
±2, 4, 8 kV air
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
3 A/m
m
80 MHz to 800 MHz
,
0.12
0.37
1.1
3.7
12
0295-001-072 Rev-1
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Power frequency magnetic fields
should be at levels characteristics of a
typical location in a typical commercial
or hospital environment.
800 MHz to 2.5 GHz
,
0.23
0.74
2.3
7.4
23
www.c2dxinc.com