4. TABLE MOVEMENTS (NORMAL AND REVERSE ORIENTATIONS)
4.1
Seat and Back Sections
4.2
Lateral Tilt
4.3
Height and Sliding
4.4
Head Section
5. TABLE RIGIDITY
5.1
Check both ends of tabletop for any horizontal or vertical play.
5.2
Check side tilt mechanism for any play.
5.3
Check central column for any play.
5.4
Check for wobbling of table base due to uneven floor locks.
Note:
Before the base cover is closed, please make sure tools are left outside of the table
base.
Remarque: Avant de refermer le couvercle de la base, assurez-vous de n'oublier aucun outil
dans la base de la table.
6.7
Battery Charging Procedures
Batteries will be charged automatically as long as the AC power is plugged in (#4, Figure 5-3).
Battery Level Indicator indicates low or discharged battery conditions on the Hand Pendant (#16,
Figure 5-2) as well as on the Base Panel (#8, Figure5-3). We recommend that AC power be
plugged in whenever possible to allow charging the batteries, since this will prolong battery's life.
Follow the battery charging schedule below, if AC power is not the main power supply:
1.
Every 10 days when the table is in normal service; more often if usage demands.
2.
Charge whenever any of the Battery Power icons indicate low or the base panel Battery
Level Indicator displays less than two green lights.
3.
If a table remains in extended storage for more than 1 month; batteries must be fully charged
before storing and charged every 1 month.
Note:
1.
For the sealed lead battery, allow about 12 hours for full battery charge. Charging time
for LiFePO4 batteries is 5 hours.
ITEMS
6-7
B2AV800 Series Surgical Table
IM-B2AV800-E01
MINIMUM FREQUENCY
1× per year
1× per year