3.6.7
Duty Cycle
The table is not designed for continuous use over an extended period of time. The duty cycle is
50%. It is recommended that after operating the table continuously for five minutes, it should
rest for about five minutes before being used again.
3.7
Electromagnetic Interference
INTERFERENCE HAZARD: Do not place this table near other equipment that will
generate electromagnetic or other types of interference.
RISQUE D'INTERFÉRENCE: N'installez pas cette table à proximité d'autres équipements
qui génèrent des interférences électromagnétiques ou d'autres types d'interférence.
B2AV800 Series surgical table meets EN 60601-1-2, and ENV50140 standards:
Medical Electrical Equipment: UL 60601-1 1st Edition, Rev. April 26, 2006 & CAN/CSA
C22.2 NO 601.1-M90, Rev. February 1998 & IEC 60601-1:1988 + A1:1991 + A2:1995
& IEC 60601-2-46: 1998
EMI regulation: CISPR 11 Group I Class A.
ESD regulation: Air discharge → 2~8kV, Contact discharge → 4kV
RF immunity regulation: 80~1000MHz, 10V/M, 1kHz, 80% AM
RF immunity regulation: 1G~2.5GHz , 10V / M , 1kHz , 100% AM (ENV50140)
EFT/Bursts regulation: ±2KV on AC and I/O ports
Surge regulation: ±2KV Common mode, ±1KV Differential mode
Conducted immunity regulation: 150kHz~80MHz, 10V(rms), 1kHz, 80% AM
Magnetic immunity regulation: 3A/m
Voltage Dips regulation: 30%/10mS, 60%/100mS, at 230VAC
3.8
Accessories
NUVO offers a wide range of accessories for its B2AV800 Series surgical tables. Please contact
an NUVO sales representative for more information about B2AV800 Series surgical table
accessories.
WARNING! AVERTISSEMENT!
3-7
B2AV800 Series Surgical Table
IM-B2AV800-E01