1.
Fuse (neutral)
2.
Fuse (live)
3.
Wire Collect Tie
4.
Power Cord Socket
5.
Manual Rotary Knob
6.
Foot Control Socket
7.
AC Power Indicator (Green LED)
8.
Battery Level Indicator(10 bars: Full, 5
bars: Mid, 2 bars: Low)
9.
(not used)
10.
ON/OFF Switch (to actuate movement)
11.
Override Control Switch (to control
direction of movement)
The B2AV800 Series tables are equipped with Override System (Figure 5-3) for controlling
table movements in the event of Hand Pendant malfunction. The system consists of an
on/off switch, a manual rotary knob and an override control toggle switch.
1.
Turn knob (# 5, Figure 5-3) to select a desired movement.
2.
Toggle switch (# 11, Figure 5-3) up or down towards the direction of desired movement.
Note:
If the table moves by itself, return the Override Control toggle switch to its
neutral position, which is in the middle, and call for service
Remarque: Si la table se déplace d'elle-même alors que vous activez le système de
commande de priorité, désactivez-le immédiatement et appelez un
technicien en entretien et en
3.
Toggle switch (#10, Figure 5-3) to activate the movement.
4.
Repeat step 1, 2 and 3 as many time as needed to create a desired table position.
Note:
 Override System is only suitable for NORMAL patient orientation.
 Although table positioning can be performed with the Override System, the problem
should be corrected as soon as possible. Have a qualified service technician repair
the table before further use.
Rotary Knob (#5)
No.
Description
12.
Floor Lock
13.
Slide
14.
Height
15.
Tilt
16.
Trendelenburg
17.
Back
réparation.
5-17
B2AV800 Series Surgical Table
IM-B2AV800-E01
Override Control (#11)
Up
Down
Lock
Unlock
To head
To leg
Up
Down
Left
Right