Monthly Maintenance; Quarterly Maintenance - Nuvo B2AV800-1 Manuel

Table des Matières

Publicité

Turn the table's power off with Hand Pendant when finished with cleaning procedure.
3.
6.3

Monthly Maintenance

Perform all table functions. Observe if all movements are smooth and if there is any noise.
Have a qualified technician check the table for anything out of the ordinary. Never allow
inexperienced, unqualified technicians to attempt any repair on the table.
6.4

Quarterly Maintenance

SHOCK HAZARD: Monitor the electrically conductive properties of the floor locks (all
models) and casters (B2AV800S(K) and B2AV800(K)). Measure their resistance level
according to the requirements of NFPA 99 before placing table in use. Subsequent
routine testing should be performed at least once every three months.
RISQUE DE DÉCHARGE ÉLECTRIQUE: Surveillez les propriétés de conductivité
électrique des dispositifs de verrouillage au sol (pour tous les modèles) et des roulettes
(pour les modèles B2AV800S(K) et B2AV800(K)). Mesurez leur niveau de résistance
selon les exigences de la norme NFPA 99 avant de commencer à utiliser la table. Des
tests de routine doivent être effectués au moins une fois tous les trois mois.
ELECTRIC SHOCK HAZARD: Only authorized service personnel should be allowed to
remove the covers.
RISQUE DE DÉCHARGE ÉLECTRIQUE: Seul le personnel de service autorisé devrait
être autorisé à retirer les couvercles.
1.
Clean the floor locks according to 6.2.2 and then check for conductivity.
WARNING! AVERTISSEMENT!
6-5
B2AV800 Series Surgical Table
IM-B2AV800-E01

Publicité

Table des Matières
loading

Table des Matières