Route hand pendant cord and optional foot control cord clear of any pinch points where
they could be damaged.
To avoid damage to the hang hand pendant, clip from side rail of table when not in use.
IMPORTANT: Operate table through all articulations prior to usage when table has been
stored for longer than four weeks,
Plug power cord into a properly grounded socket only.
Check the manual and be familiar with all information before operation.
When using a third party arm support, keep an angle greater than 15 degrees along the
direction of table length to avoid component damage during tabletop movements.
Do not articulate table into any angle that could hurt the patient.
Do not place this surgical table near other equipment that will generate electromagnetic
interference or other types of interference.
Before using high-frequency surgical equipments, cardiac defibrillators or cardiac
defibrillator-monitors, please refer to their instruction manual.
Use only NUVO's mattress pads in order to avoid risk of electrostatic shock.
When the power cord is damaged in any way, replace it immediately or else use the
table's internal electrical power source until it can be replaced.
The antistatic pathway is: Mattress Pad→Tabletop→Column→Base→Floor Lock
Pads→Floor. Therefore, the anti-static properties of the table depend upon the use of
the recommended NUVO mattress pads.
If the surgical drape hangs over the articulation points or the articulation covers (stainless
steel plates covering the joints), movement could cause damage to the table. Tuck the
drapes between the mattress pad and tabletop in order to prevent this.
There is only one way to insert the connector of the hand pendant into its receptacle on
the table. The connector must be inserted straight into the receptacle without twisting it,
CAUTION!
5-5
B2AV800 Series Surgical Table
IM-B2AV800-E01