5.6.2
Distribution tap calibration
You must calibrate the distribution tap if:
• The tap motor has been replaced
• The error message: <Tap Calibration> appears on the display.
Carry out the calibration as follows:
Fig. 24:
Calibration of the distribution tap
1. Switching on the device
2. Under <SET -> SERVICE - CALIBRATION>, select the
<DISTR. TAP> menu item.
3. By accepting the "start" option, the tap moves and stops just
before the calibration position.
4. At the controls, keep pressing "1 step" until the arrow on the
front of the tap is precisely in the recess in the center of the
distribution pan (Point A). From the picklist, then select the
<SAVE> menu item.
5. The tap is calibrated.
Operation