Commissioning
Parameter
Set
Multi-probe
Basic settings
Multi-probe
Set
Memory interval
Basic settings
Memory Interval
Amount In
Set
Basic settings
Memory Interval
Capacity
Set
Basic settings
Memory Interval
Reset
Set
Basic settings
Memory Interval
Mode
Set
Interfaces
Basic settings
Interfaces
Program selection
Set
Number
Program selection
Number
Set
Prog. nr.
Program selection
Prog. nr.
Set
Switch (only for an
active switching
Program selection
program)
Switch Time
Set
Program selection
Switch Ext.
Set
Program selection
Switch Flow
116
Menu level
Function to activate, deactivate and configure the multi-probe, and to
read out measured values.
Time interval for internal recording of measured flow values. 0 sec =
no recording.
The capacity of the ring memory (in hours) is shown automatically on
the display.
Activate/deactivate ring memory resetting when the Aut button is
pressed.
The data recorded for the analog input and the probe etc. can be saved
even if a program is not active. For this purpose, you must set the
"Mode" parameter to "Continuous". The device saves to the ring
memory as soon as it is in the "Unit On" mode.
Transmission rate and definition of the RS232 interface
1: 1 main program active
2: 1 main program and 1 switching program active
1+event: 1 main program and one event program active
2+event:1 main program, 1 switching program and 1 event Program
active
Selection of active sampling program
Switch.: Inactive, Active
Options: Day (switch between main program and switching program
at two specific times a day)
D Switch 1: 00:00
D Back 1: 00:00
D Switch 2: 00:00
D Back 2: 00:00
Options: Week (switch between main program and switching
program on three specific days a week)
W Switch 1: Mon, Tues, Wed, Thurs, Fri, Sat, Sun, Off
W Switch 2: Mon, Tues, Wed, Thurs, Fri, Sat, Sun, Off
W Switch 1: Mon, Tues, Wed, Thurs, Fri, Sat, Sun, Off
Switching on external signal (assign signal input!).
Switch.: Inactive, Active
Switch Pr.: Inactive, Active
Hysteresis: 0001
Up. Setp: Inactive, Active (switch if flow is larger than limit value)
Lo. Setp.: Inactive, Active (switch if flow is smaller than limit value)
Setting options
ASP station 2000 peristaltic
Factory setting
Off
0000 sec
Aut Button
Prog. Active
1
Program 1-4;
1-U1; 1-E;
1-U1-E
Inactive
Day
00:00
00:00
00:00
00:00
Off
Off
Off
Inactive
Inactive
0001 l/s
Inactive
Inactive