Endress+Hauser
The programme has been started
Fig. 13:
"Off" key, active running programme
If a programme is running, pressing the "Off" key briefly (<= approx. 1 s) sets the programme to
break mode.
If the key is pressed for a longer period (> 1 s), it terminates the running programme.
Fig. 14:
"Off" key, break mode
In the break mode, the user can decide whether the programme should be terminated. The
programme is terminated by pressing the "Off" key again. If the programme should be resumed, the
programme is resumed once the "On" key is pressed. Here, the tap goes through its reference
position once and then on to the current bottle. Any bottle changes which are due to take place
during the break period are performed.
5.1.3
"Rep" key
Fig. 15:
"Rep" key
Press the "Rep" key to activate the display of the report list. Events are logged in this list while the
device is switched on.
A maximum of 30 reports are stored. The reports are stored in a ring memory. If the ring memory
is full and a new report occurs, the oldest report is deleted.
Report
Unit on
Reports on the display
Operation
Report list
14:12 11.01.98