Failure / Malfunction
Failure / Malfunction
If you press the manual control push-button no pulse is conveyed to the bone-
If you press the manual control push-button no pulse is conveyed to the bone-
expander/osteotome.
expander/osteotome.
Possible cause
Possible cause
The manual control push-button is damaged.
The manual control push-button is damaged.
Internal circuit failure
Internal circuit failure
Possible solution
Possible solution
Verify if the pulse is conveyed by pressing the footswitch. If so contact the
Verify if the pulse is conveyed by pressing the footswitch. If so contact the
authorized service asking for the manual control switch repair/replacement.
authorized service asking for the manual control switch repair/replacement.
Contact the authorized service.
Contact the authorized service.
Failure / Malfunction
Failure / Malfunction
The display is switched off or it shows some nondescript number.
The display is switched off or it shows some nondescript number.
Possible cause
Possible cause
The display is damaged.
The display is damaged.
Internal circuit failure.
Internal circuit failure.
Possible solution
Possible solution
Take note that the level of the force is all the same indicated by the knob po-
Take note that the level of the force is all the same indicated by the knob po-
sitioning (1-2-3-4). In that event you can follow through the treatment before
sitioning (1-2-3-4). In that event you can follow through the treatment before
calling the authorized service for display replacement or repairs.
calling the authorized service for display replacement or repairs.
Failure / Malfunction
Failure / Malfunction
Mild heating of the handpiece
Mild heating of the handpiece
Possible cause
Possible cause
Improper use. You have exceeded the maximum threshold of 200 consecutive
Improper use. You have exceeded the maximum threshold of 200 consecutive
pulses without having had the necessary break.
pulses without having had the necessary break.
Possible solution
Possible solution
Switch off the device. Wait 40 minutes before switching on the device again.
Switch off the device. Wait 40 minutes before switching on the device again.
Failure / Malfunction
Failure / Malfunction
The o-ring type 108 of the handpiece is no more working as anti-unscrewing for
The o-ring type 108 of the handpiece is no more working as anti-unscrewing for
the ring nut
the ring nut
Possible cause
Possible cause
O-ring consumption or breaking
O-ring consumption or breaking
Possible solution:
Possible solution:
Replace the O-ring with the spare one supplied with the handpiece (see
Replace the O-ring with the spare one supplied with the handpiece (see
blades container).
blades container).
If the problem endures, contact the authorized service provider.
If the problem endures, contact the authorized service provider.