• Disconnect alarm
This alarm alerts the caregiver that the Remote Alarm has been
disconnected from the Ventilator.
Check the patient and ventilator prior to looking for the source of
the disconnect.
Note: For further information regarding alarm types and
indicators refer to the Indicators and alerts section in this Guide.
Indicators and alerts
The Remote Alarm provides the following indicators and buttons.
LED indicators
Alarm
Volume
Disconnect
Battery level
Alarm button
4
Audible alert and
Red flashing alarm
bar.
Blue solid LED
Audible alert and
Yellow blinking
LED.
Audible alert and
Yellow LED – will
blink once every
10 seconds.
Green LED – will
blink once every
10 seconds
Volume level
button
Remote Alarm
mute button
Remote Alarm
self-test.
Status
An alarm has been activated
on the ventilator.
Indicates current alarm
volume level.
• The Remote Alarm has
been disconnected from
the ventilator.
• The LED will continue to
blink until the situation has
been resolved.
• The Remote Alarm battery
is low.
• The LED will continue to
blink until the situation has
been resolved.
• The audible alarm can be
muted but will sound every
4 hours until the battery is
replaced.
The Remote Alarm battery
is good.
Press until desired volume
level is achieved. Refer
to Setting the volume
instructions.
Press the Remote Alarm mute
to silence the alarm generated
for a Low Battery alert.
Pressing the Remote Alarm
mute during a disconnect will
put the Remote Alarm into a
power saving mode.
Press and hold to test the
functionality of the Remote
Alarm.