Alarm Volume Range
IEC 60601-1-11:2020
classification
Nominal dimensions
(L x W x H)
Power supply
Weight
Housing construction
Cleaning solutions
Compatible devices
Maximum service life
Symbols
The following symbols may appear on your product or packaging.
Indicates a Warning or Caution. A warning alerts you to possible
injury or explains special measures for the safe and effective use of the
product; a caution alerts you to possible damage to the product;
instructions before use;
The lower audio level will generate a
volume: 65 dBa +/- 5 dBa
The medium audio level will
generate a volume: 75 dBa +/- 5 dBa
The higher audio level will generate
a volume: 85 dBa +/- 5 dBa
Transit operable equipment.
145mm x 80mm x 31mm
Two AA 1.5V Batteries
(LR6 – Alkaline or FR6 – Lithium)
115g Remote Alarm only (without
battery)
Flame retardant engineering
thermoplastic
The following cleaners and
disinfectants are compatible for use
when cleaning the external surfaces
of the Remote Alarm. They are
suitable for weekly use.
• Cavicide
• Mikrozid
• Isopropyl alcohol
• Teepol Multipurpose detergent
and Actichlor Plus
• Bleach (1:10) (dilute Hypochlorite)
* Suitable for monthly use only.
Always follow the manufacturers
recommended cleaning instructions.
The Remote Alarm II is compatible
with the following ResMed devices:
• Stellar series of ventilators
• Astral series of ventilators.
• When the Remote Alarm is
connected to the ventilator, and
assuming that the ventilator
does not alarm more than 1% of
the time, the battery life of the
Remote Alarm shall be more than
6 weeks (ambient temperature
21°C and ambient humidity not
condensing)
• When the Remote Alarm is
connected to the ventilator
in standby (not alarming), the
battery life of the ventilator shall
be more than 12 weeks (ambient
temperature 21°C and ambient
humidity not condensing)
Keep away from rain;
*
™
AF*
®
Drip proof;
IP22
Read
English 7