4 Intended use
4.1 Indications for use
System ankle joints are intended to be used exclusively for providing a
patient with a lower limb orthosis.
In case of unilateral use, the manufacturer recommends that the system
ankle joint be processed by means of the lamination resin technique or car
bon prepreg technology.
4.2 Indications
Partial or total paralysis of the leg muscles.
Indications must be determined by the physician.
4.3 Combination possibilities
Combination 17AD100* with 17B203* E-MAG Active
The 17AD100* Nexgear Tango orthotic ankle joint can be combined with the
17B203* E-MAG Active orthotic knee joint.
Orthotic knee
joint
17B203*
17B203* with
17B206 medial
support
INFORMATION
Do not make any further static or dynamic changes to the 17AD100*
after automatic calibration of the 17B203* E-MAG Active.
If patient-specific adjustments to the 17AD100* are required, repeat the
process steps for automatic calibration of the 17B203* E-MAG Active
according to the 647G1165 instructions for use.
Further information about the processing of the 17B203* E-MAG Active is
found in the 647G1165 instructions for use.
4.4 Lifetime
The product is designed for the following lifetime, given intended use and
proper installation:
Joint sizes 10, 12: 1 year
34
Combination possibilities
Max. body
weight [kg]
(lbs)
85 (187.4)
100 (220.5)
Orthotic ankle
joint
17AD100=16*,
=20*
17AD100=16*,
=20*
Usage condi
tion
Unilateral
Bilateral