Sunrise Medical Oxford Ascend Manuel De L'utilisateur Et Garantie page 14

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

14
4. Emergency Stop
The red Emergency Stop Button is located on the front of the control box and is acti-
vated by pressing in. This will cut all power to the lift and only be reset by twisting
the button anticlockwise and releasing.
5. Electrical Emergency Down/Up
Emergency lower (down) and raise (up) buttons are provided at the front of the control
box. This is operated by inserting a ballpoint pen into the button highlighted
This will bring the boom up or down should the handcontrol fail at any time.
6. Mechanical Emergency Down
In the case of a complete electrical failure the electrical actuator is fitted with mechan-
ical lowering device (RED BOSS). This will only operate when the lift is under load. The
device must be pulled upwards to activate, and a slow decent will commence.
7. Batteries
The batteries are protected from deep discharge by a LOW VOLTAGE ALARM. This will
sound when the batteries need recharging and the handcontrol is being operated. It
will not sound independently of the handcontrol being operated. DO NOT IGNORE THIS
WARNING ALARM. Complete the lifting operation and place the battery on charge (see
charging instructions).
8. Slings
The Oxford/Hoyer Ascend is suitable for patients in the SITTING, SITTING/RECUMBENT
and RECUMBENT positions. The slings suitable for this device are listed as follows:
* Oxford/Hoyer Transport sling
* Oxford/Hoyer Standing sling
9. Knee pad adjustment
The Ascend has an adjustable kneepad. The range of travel is 6 inches. Adjustment is
achieved by loosening the knob located on the left hand side of the pad, so to loosen
the assembly enough to move the knee pad in either the upward, or downward position.
10. Removable footplate
The Ascend has a removable footplate. To remove the foot plate, move to the front of
the product, and lift the footplate upward from the middle of the assemble. NOTE Be
careful when lifting the footplate from its low-seated position.