a The fixed transmitter field levels, such as the base stations of the radio telephones (cellular/wireless) and
the terrestrial mobile radios, domestic radio, AM, FM, and TV radio communication cannot be theoretically
assessed precisely. To obtain the electromagnetic environment due to the fixed RF transmitters, a site
measurement should be performed. If a field level measured in the use environment of SOPROLIFE device
exceeds the compliance levels above applicable, the good operation of SOPROLIFE device should be
checked. If abnormal operations are proved, some further measures should be taken, such as reorientation
or relocation of the standard device.
b Above the 150 kHz to 80 MHz frequency band, the field level should be lower than 3 V/m.
Recommended separation distances between the portable and mobile RF communication devices
SOPROLIFE device is intended to be used in an electromagnetic environment in which the irradiated RF
disturbances are checked. The user of SOPROLIFE device can help to avoid electromagnetic interference by
maintaining a minimal distance between the portable and mobile RF communication devices (transmitters)
and the recommended SOPROLIFE device such as recommended below, depending on the maximum output
power of the communication device.
Rated maximal output
power of the
transmitter
W
0.01
0.1
1
10
100
For the transmitters whose maximal output is not listed above, the recommended separation distance d in
metres (m) can be determined by using the equation applicable to the transmitter frequency, where P is the
maximal output of the transmitter in watts (W) rated by the transmitter manufacturer.
Note 1: At 80 MHz and at 800 MHz, the separation distance given in the higher frequency band applies.
Note 2: These recommendations may not apply in every situation. The electromagnetic wave propagation is
modified by absorption and reflection due to the structures, objects and persons.
10
S O P R O L I F E • U s e r ' s m a n u a l
E N G L I S H
and SOPRO device
Separation distance depending on the transmitter frequency
150 kHz to 80 MHz
d = 1,16 √P
0.116
0.366
1.16
3.66
11.6
m
80 MHz to 800 MHz
d = 1,16 √P
0.116
0.366
1.16
3.66
11.6
800 MHz to 2,5 GHz
2.33 √P
0.233
0.736
2.33
7.36
23.3