Guide and declaration of the manufacturer - electromagnetic immunity
SOPROLIFE device is intended to be used in the electromagnetic environment specified below.
Immunity trial
Electrostatic discharges
EN 61000-4-2
Far transient bursts
EN 61000-4-4
Voltage shocks
EN 61000-4-5
Dips, brief outages and
power voltage variation
EN 61000-4-11
Magnetic field with the
network frequency
(50/60 Hz)
Note: U
is the power voltage nominal value applied during the trial.
T
8
S O P R O L I F E • U s e r ' s m a n u a l
E N G L I S H
The user should make sure it is used in this environment.
CEI 60601
Compliance
Severity level
level
± 6 kV when
± 6 kV
in contact
± 8 kV in the
± 8 kV
air
± 2 kV for the
± 2 kV
feed cables
± 1 kV for the
± 1 kV
input/output
cables
Differential
± 1 kV
mode ± 1 kV
Common
N.A.
mode ± 2 kV
• <5% U
- for
<5% U
T
10 ms
10 ms
• 40% U
- for
<40% U
T
100 ms
100 ms
• 70% U
- for
<70% U
T
500 ms
500 ms
• <5% U
- for
<5% U
T
5 s
5 s
3 A/m
3 A/m
Electromagnetic environment
The floor should be wooden, concrete or tile. If
the floor is covered with a synthetic material, the
relative humidity should be at least 30%.
The main power supply quality should be one of a
traditional commercial or hospital environment.
The main power supply quality should be one of a
traditional commercial or hospital environment.
The main power supply quality should be one of a
T
traditional commercial or hospital environment.
If the user of SOPROLIFE device requires it to
T
continue to operate during main power supply
outages, it is recommended SOPROLIFE device is
T
fed by an inverter or a battery.
T
The magnetic field with the network frequency
should be at a characteristic level of a location in
a traditional commercial or hospital environment.