8.9

Interruption of sterilization cycle

A sterilization cycle can be voluntarily interrupted by pressing the button
least 2 seconds. The autoclave will emit a beep, go into decompression and on the display
will appear the message ALLARM 001 (INTERRUPTED CYCLE).
To reset error, maintain pressed simultaneously the buttons below indicated, until the
intervention of an acoustic signal:
ATTENTION: You can't reset the alarm if the printer works
ENGLISH
B-START
for at
B-SELECT
B-PUMP
B-DOOR